What Is Particle Size Distribution? D10, D50, D90, and Measurement Methods
12 min read••By Chemical Express Thailand

Your batch release stalls because your D50 reading does not match the supplier certificate, and procurement wants to know which number to trust. The answer is often neither, because sampling protocol, distribution basis, and instrument choice each pull results in different directions, even when the powder itself has not changed. This article explains what particle size distribution means, how D10, D50, and D90 are defined and measured, which technique fits which workflow, and how to report results that hold up under QA review.
What is particle size distribution and what do D10, D50, D90 mean?
Particle size distribution is the statistical description of particle diameters across a sample population, reported as a cumulative curve or a set of percentile values read from that curve. D10, D50, and D90 are the diameters at which 10%, 50%, and 90% of the measured population fall below that size, respectively. They compress a full size profile into three actionable numbers.
The critical detail most labs overlook: D-values depend entirely on the distribution basis selected. Laser diffraction instruments typically report volume-based D10, D50, and D90. Dynamic light scattering instruments report intensity-weighted and number-weighted sizes by default. A volume-based D50 of 8 µm and a number-based D50 of 2 µm can describe the same powder without contradiction.
| Parameter | Definition | Typical Basis | Practical Use |
|---|---|---|---|
| D10 | 10% of particles are below this diameter | Volume (laser diffraction); number or intensity (DLS) | Characterises fine fraction; relevant to dissolution, packing voids |
| D50 | 50% of particles are below this diameter (median) | Volume, number, or intensity depending on method | Primary QC acceptance criterion; compares batches |
| D90 | 90% of particles are below this diameter | Volume (laser diffraction); intensity (DLS) | Controls coarse tail; critical for filter sizing, inhalation, electrode density |
| Span | (D90 − D10) ÷ D50 | Same basis as D-values used | Single-number width indicator; lower span means narrower distribution |
How PSD is reported and interpreted: basis, curves, and spread
Cumulative vs. differential curves
The cumulative distribution plots the fraction of particles below a given size against that size. You read D10, D50, and D90 directly off the y-axis at 10%, 50%, and 90% marks. The differential (or frequency) curve is the derivative of the cumulative curve and shows where particle counts or volumes concentrate, making modes and shoulders in a bimodal sample visible at a glance.
Always inspect the differential curve before accepting a D50 number. A bimodal sample with a large fines population and a coarse agglomerate fraction can yield the same D50 as a narrow monomodal powder, yet behave very differently in process.
Distribution weighting: number, volume, mass, and intensity
Number weighting counts each particle equally, so a few large particles barely register. Volume (or mass) weighting scales each particle by its volume, meaning a single 100 µm sphere outweighs a million 1 µm spheres. For irregular shapes, instruments assume a spherical equivalent diameter.
A worked example makes this concrete. Imagine a bimodal mix of mostly 1 µm particles with a small fraction of 50 µm particles. The number-based D50 sits near 1 µm because small particles dominate by count. The volume-based D50 shifts toward 40 µm or higher because a few large particles dominate the volume. Both values are correct; they answer different questions.
In DLS, intensity-weighted results emphasise large or highly scattering particles even more strongly than volume weighting does. Converting between intensity, volume, and number distributions requires refractive index assumptions and introduces compounding uncertainty, so always report the native output alongside the distribution basis.
Spread descriptors and dispersion indicators
Span = (D90 − D10) ÷ D50 is the standard breadth descriptor for laser diffraction data. A span below 1.5 is generally considered narrow for pharmaceutical powders; values above 2.5 often signal broad or multimodal distributions. For DLS data, the polydispersity index (PDI) serves a similar role: PDI below 0.1 indicates a narrow, monomodal dispersion; above 0.5 suggests that mean diameter values are unreliable.
Quick tips: sampling and dispersion
- Use a riffle splitter or coning-and-quartering to obtain a representative sub-sample from bulk powders.
- Set ultrasonication energy conservatively: over-sonication breaks primary particles and artificially shifts D10 downward.
- Add an appropriate surfactant or dispersant to prevent re-agglomeration, especially for hydrophobic powders in aqueous media.
- Pre-sieve to remove oversize agglomerates that would otherwise dominate volume-weighted D90.
- Verify obscuration (laser diffraction) or count rate (DLS) sits within the instrument's validated window before recording data.
- Run at least triplicates per sample; report mean, standard deviation, and percentage relative standard deviation.
Particle size distribution measurement methods and how they work
The right technique depends on particle size range, sample concentration, optical properties, and the distribution statistic your specification requires. No single method is universally best. Matching instrument physics to sample reality saves time in method development and prevents out-of-specification investigations rooted in technique mismatch rather than real product variation.
Core methods
Laser diffraction covers a wide dynamic range, typically from below 100 nm to several millimetres, in a single run. The instrument applies Mie or Fraunhofer scattering models to angle-resolved intensity patterns and reports volume-based D-values. Refractive index input and dispersion homogeneity directly control the accuracy of D10 and D90, where the scattering signal is weakest.
Dynamic light scattering (DLS) measures fluctuations in scattered laser intensity caused by Brownian motion and calculates the hydrodynamic diameter. It works best for particles in the 1 to 1000 nm range in dilute, optically transparent dispersions. DLS reports intensity-weighted size as its primary output; even a small number of aggregates can dominate the intensity signal and inflate apparent size.
Sieve analysis is robust and mass-based, covering approximately 20 µm to several millimetres. Many pharmacopeial and industrial material specifications still reference sieve fractions directly. Resolution is limited, and elongated or flaky particles pass through openings larger than their minimum dimension, introducing a shape bias that is absent from ensemble optical methods.
Image analysis and optical microscopy provide direct size measurements and shape descriptors such as aspect ratio and circularity. These methods confirm multimodality, identify agglomerates, and characterise non-spherical particles in ways that ensemble methods cannot. Throughput is lower, but image analysis increasingly automates particle counting on thousands of particles per run.
Sedimentation and centrifugal methods use Stokes' law to relate settling velocity to particle diameter in the submicron to tens-of-micrometres range. They require accurate density values and stable, non-flocculating dispersions. Centrifugal disc sedimentation offers good resolution for narrow distributions but is less common in routine QC.
Method selection factors
Match your technique to the expected size range, sample physical state, and required output statistic first. Then consider refractive index availability, opacity, and expected polydispersity. For regulated products, check whether your pharmacopeia or product standard cites a preferred method or allows alternatives with validation.
Quick decision guide: laser diffraction vs. DLS vs. sieve analysis
- Submicron to millimetre range in one run
- Volume-based D10, D50, D90 needed for spec
- Wet or dry dispersed powders and slurries
- High-throughput QC with reproducible dispersion
- Nanoparticles in the 1 to 1000 nm range
- Dilute, transparent colloidal dispersions
- Stability monitoring via PDI and Z-average tracking
- Hydrodynamic diameter for biological or polymer nanoparticles
- Coarse powders from 20 µm to several millimetres
- Mass-based fractions required by specification
- Pharmacopeial or industrial standards that reference mesh sizes
- Simple, low-cost routine checks on granules or pellets
| Technique | Size Range | Sample Type | Output Basis | Throughput | Common Failure Modes |
|---|---|---|---|---|---|
| Laser diffraction | ~0.1 µm to ~3.5 mm | Dry powder, wet dispersion, spray | Volume | High | Wrong refractive index; poor dispersion; multiple scattering in dense samples |
| Dynamic light scattering | ~1 nm to ~1 µm | Dilute liquid dispersion | Intensity (primary); volume and number derived | High | Aggregates dominate signal; dust contamination; sample too concentrated |
| Sieve analysis | ~20 µm to several mm | Dry or wet powder/granule | Mass | Moderate | Blinding; shape bias; electrostatic charging; limited resolution |
| Image analysis | ~0.5 µm to several mm | Dispersed particles or powder feed | Number (area-equivalent or Feret) | Low to moderate | Low statistics on rare large particles; sample prep artefacts |
| Sedimentation / centrifugal | ~0.05 µm to ~100 µm | Stable liquid dispersion | Mass (Stokes equivalent) | Low to moderate | Flocculation during run; density uncertainty; convection currents |
Consumables, reference materials, and synthesis reagents used in sample preparation and formulation workflows can be sourced from Merck Materials Science through Chemical Express Thailand, helping keep your characterisation workflow consistent from raw material to measured result.
Applications and specifications: pharma, food, ceramics, batteries, and catalysts
Pharmaceutical solids and inhalation
Dissolution rate and bioavailability of oral solid dosage forms correlate strongly with D50 and span. A smaller D50 increases surface area and speeds dissolution, but a high span indicates fines that can cause segregation and content uniformity failures. Inhalation powders require a tight respirable fraction: the entire D10 to D90 window typically must fall within 1 to 5 µm for deep lung deposition.
Food, nutraceuticals, and emulsions
Texture and mouthfeel in food systems depend on whether particles remain below the sensory detection threshold, typically around 30 µm. Emulsion and suspension stability correlates with D50 and the D90 tail: a rising D90 over storage time signals coalescence or flocculation before visible phase separation occurs. Tracking D10, D50, D90 at multiple time points gives early warning on shelf-life performance.
Ceramics and advanced materials
Packing density and sintering behaviour in ceramics depend on the full PSD shape, not just the median. Multimodal blends, where coarse particles fill the framework and fine particles occupy voids, maximise green density. The rheology of ceramic slurries is particularly sensitive to fines content in the low end of the distribution, which drives viscosity and affects tape casting or slip casting outcomes.
Batteries and energy materials
Cathode and anode active materials use tight PSD control to manage electrode packing density, porosity, and rate capability. A D50 that is too coarse reduces accessible surface area; a D90 with large agglomerates creates local hotspots and limits fast-charge performance. Conductive additives and binders must themselves be free of oversize agglomerates to coat uniformly.
TCI Material Building Blocks provide feedstocks and molecular precursors used in materials synthesis workflows where downstream PSD control starts at the synthesis stage. Using well-characterised building blocks reduces batch-to-batch variability in particle formation steps.
Catalysts and adsorbents
External surface area and mass transfer resistance in packed-bed catalysts are direct functions of particle size. A tight PSD reduces pressure drop variation and improves bed uniformity. The coarse tail of the distribution also affects attrition resistance: large particles fracture preferentially under mechanical stress, generating fines that cause pressure drop to rise over the catalyst lifetime.
| Application | Typical Acceptance Criterion | Preferred Method |
|---|---|---|
| API (oral solid dosage) | Volume-based D50: 5 to 10 µm; Span < 1.8 | Laser diffraction (wet) |
| Spray-dried powder | D90 < 50 µm | Laser diffraction (dry or wet) |
| Inhalation dry powder | D10 > 1 µm; D90 < 5 µm | Laser diffraction (wet, low-shear) |
| Li-ion cathode material | Volume-based D50: 6 to 12 µm | Laser diffraction (wet) |
| Ceramic slurry | D90 < 10 µm; no oversize particles | Laser diffraction (wet) |
| Pharmaceutical granules | Sieve fraction passing 500 µm > 95% | Sieve analysis |
For researchers sourcing reference materials, synthesis chemicals, or materials aligned to specific application workflows, the Materials Science product range at Chemical Express Thailand covers categories from nanoparticles to energy materials and biomedical polymers.
Summary and next steps for lab teams
Particle size distribution describes how particle diameters spread across a sample population, and D10, D50, D90 are percentiles that only carry meaning when the distribution basis and measurement method are explicitly stated. Switching basis or technique without documentation is the most common source of inter-lab and supplier vs. customer disagreements.
Adopt a method that fits your size range and sample matrix, write the dispersion protocol and refractive index inputs into a validated SOP, and run triplicates as a minimum. Use the comparison table above when deciding between laser diffraction, DLS, and sieve analysis. If your distribution is multimodal, inspect the differential curve and report additional percentiles or mode positions alongside D10, D50, and D90.
Review available materials categories to align your sample preparation and synthesis reagent sourcing with your characterisation requirements. Clear, documented specifications protect both your QC decisions and your supplier negotiations.
Frequently Asked Questions