What Is Viscosity? How Resistance to Flow Is Measured

A QC batch is trending thick, and the team must decide whether to adjust solvent ratio or hold the lot. Before that call, someone needs a confirmed viscosity reading at the right temperature and shear conditions, with a method the SOP actually permits. This article defines viscosity, explains the underlying physics, compares measurement techniques, and provides selection tables and setup tips to get numbers you can trust.

What is viscosity in simple terms?

Viscosity is a fluid's internal friction that resists flow. Dynamic viscosity describes resistance under an applied shear stress, while kinematic viscosity describes how a fluid flows under gravity and equals dynamic viscosity divided by density at the same temperature.

Units matter and must always be paired with a temperature. Dynamic viscosity is reported in Pa·s or mPa·s (equivalent to centipoise, cP). Kinematic viscosity is reported in m²/s or mm²/s (equivalent to centistokes, cSt). For non-Newtonian samples, also state the shear rate.

Labs measure viscosity using capillary viscometers for Newtonian liquids at low to medium viscosities, and rotational viscometers or rheometers for a wider range that includes non-Newtonian fluids. In both cases, tight temperature control and documented method parameters are what make results repeatable and comparable.

How viscosity behaves under shear and temperature

Dynamic viscosity relates shear stress to shear rate. Increase the shear force and you observe how the fluid responds. For Newtonian liquids, the ratio stays constant regardless of shear rate. For non-Newtonian materials, that ratio changes, giving an apparent viscosity that depends on shear rate or on how long shear has been applied.

SymbolQuantitySI UnitCommon Lab Unit
ηDynamic viscosityPa·smPa·s (cP)
νKinematic viscositym²/smm²/s (cSt)
ρDensitykg/m³g/cm³
Temperature control: Set and hold the bath at the target temperature before loading any sample. Even a 1 °C drift can shift viscosity readings by several percent. Allow full thermal equilibration before recording the first measurement.

Newtonian vs non-Newtonian behavior

Newtonian liquids maintain constant viscosity regardless of shear rate. Common examples include water, toluene, and glycerol-water blends at low polymer content. These are the straightforward cases where a single viscosity value carries full meaning at a stated temperature.

Non-Newtonian materials are common in real-world applications. Paints, ketchup, latex, slurries, polymer solutions, and blood all show viscosity that depends on shear rate or time. Some thin as shear increases (shear thinning); others thicken (shear thickening). Some recover slowly after shear is removed (thixotropy).

For non-Newtonian materials, a single viscosity number is rarely sufficient. Report method, geometry, temperature, shear rate or spindle speed, and any time-dependent behavior. If the material is shear thinning or thickening, provide a full flow curve covering the shear rates relevant to the process.

How viscosity is measured in the lab

The two main measurement workflows in laboratory practice are kinematic (capillary-based) and dynamic (rotational or torque-based). Each suits different sample types and different decisions. Density measurement is typically required to convert between the two.

Common viscometer types

Capillary viscometers (Ubbelohde or Ostwald type) measure the time a Newtonian liquid takes to flow under gravity through a calibrated tube. Kinematic viscosity is calculated directly from flow time and the instrument constant. Convert to dynamic viscosity by multiplying by density measured at the same temperature.

Rotational viscometers and rheometers apply a controlled shear to the sample using a spindle, concentric cylinder, or cone-plate geometry, and measure the resulting torque. They cover a broad viscosity range and can characterize non-Newtonian behavior, reporting apparent viscosity at a stated shear rate. This is the method of choice when flow behavior under process conditions matters.

Falling ball viscometers infer viscosity from the terminal velocity of a calibrated ball dropping through a sample in a tube. Suitable for transparent Newtonian liquids within defined viscosity ranges. Less flexible than rotational methods but simple and reliable for qualifying oils and clear solutions.

Vibrational and cup methods (such as Ford cup or vibrating probe) give rapid screening results useful for batch consistency monitoring. They are practical at-line or in the field, but do not provide the full characterization that rotational rheometry delivers.

Which viscometer fits your sample?

Capillary (Ubbelohde / Ostwald)
  • Clear, Newtonian liquids
  • Kinematic viscosity reporting (cSt)
  • Low to medium viscosity range
  • Standardized methods (ISO, ASTM) for solvents and oils
Rotational / Rheometer
  • Non-Newtonian materials: gels, pastes, polymers
  • Apparent viscosity at defined shear rate
  • Wide viscosity range, including high-viscosity samples
  • Flow curves and time-dependent behavior
Falling Ball
  • Transparent Newtonian liquids
  • Simple verification and QC screening
  • Well-suited for oils and clear organic solvents
MethodTypical Viscosity RangeSample TypeKey Required ParametersTypical Reporting Format
Capillary (Ubbelohde)0.3 to ~10,000 mm²/sNewtonian, transparent liquidsTemperature, tube constant, density (for dynamic conversion)mm²/s (cSt) at stated °C
Rotational / Rheometer1 mPa·s to >10⁶ mPa·sNewtonian and non-NewtonianGeometry, temperature, shear rate or speed, pre-shear protocolmPa·s or Pa·s at stated shear rate and °C
Falling Ball0.5 to ~10,000 mPa·sClear Newtonian liquidsTemperature, ball calibration constantmPa·s (cP) at stated °C
Vibrational / CupMethod-dependentProcess fluids, paints, inksTemperature, cup size or probe frequencyRelative or empirical units; seconds or mPa·s

Method setup, calibration, and quality control

A consistent setup sequence prevents most repeatability problems. Follow the same steps for every run.

Quick tips: viscometer setup sequence

  • Equilibrate bath and sample together at the target temperature before starting
  • Verify instrument level and geometry alignment before each session
  • Confirm temperature at the sample point, not just the bath setpoint
  • Select shear rate or spindle speed appropriate to the sample range
  • Run a minimum of three replicates and discard outliers with cause documented
  • Clean and dry all wetted components between samples to prevent carryover
  • Record temperature, method, geometry, shear rate, and sample lot with every result

Calibration should use certified reference fluids covering low, mid, and high ranges. Track instrument response over time to catch drift early. A viscometer that passes suitability at commissioning but drifts over months will produce data that fails inter-lab comparison.

When cleaning between samples, high purity solvents help prevent residue carry-over that shifts subsequent readings. Merck's Chemistry and Biochemicals portfolio includes a broad range of lab-grade solvents and reagents suited to cleaning viscometer components and preparing calibration solutions. Using defined-grade solvents and recording the lot number used for cleaning gives you a traceable cleaning record that supports method validation.

Reporting requirement: Always wait for confirmed thermal equilibrium before recording a reading. For kinematic measurements, record density at the same temperature used for flow time. For non-Newtonian samples, state geometry, shear rate, and any pre-shear or conditioning protocol alongside the viscosity value.

Sample handling affects results more than many labs expect. Minimize evaporation between sampling and measurement, especially for volatile solvents. Degas samples gently to remove air bubbles before loading. Filter particulates only when the method explicitly permits it, and document that step clearly. Record pre-shear or mixing history for complex fluids, since mechanical history can change the apparent viscosity of thixotropic materials.

TCI's Analytical Reagents range supports the preparation of solutions and standards used in viscometry. Titration reagents, gravimetric reagents, and high-purity solvents from TCI let you prepare reference standards and cleaning solutions with well-characterized purity, which keeps your calibration traceable and reduces ambiguity in results.

CHEMEX NEWSLETTER

Get updates related to this article

Receive selected scientific articles, practical lab guides, and research insights based on your interests. NO SPAM GUARANTEED - No more than one email per week!

Please accept the Privacy Notice.

Unsubscribe at any time.

Where and when to use viscosity data in Thai labs

Viscosity measurement appears across a wide range of industries. The table below maps common application segments to the typical requirements that drive method selection.

Application SegmentTypical Viscosity RangePreferred Units and TemperatureCommon Method
Pharma and biopharma QC1 to ~10,000 mPa·smPa·s at 20 °C or 25 °CRotational (biologics); capillary (buffers, solvents)
Food and beverage10 to >100,000 mPa·smPa·s or Pa·s at process temperatureRotational; Brookfield-type
Petrochemicals and lubricants2 to 1,000+ mm²/scSt at 40 °C and 100 °CCapillary (ISO, ASTM)
Coatings, inks, and adhesives100 to >10,000 mPa·smPa·s at defined shear rate and °CRotational; cone-plate
Polymers and monomersDilute solution to bulk meltIntrinsic viscosity (dL/g) or mPa·sCapillary (intrinsic); rotational (melt)
Bioprocess and diagnostics1 to ~1,000 mPa·smPa·s at 37 °C or 25 °CRotational; micro-viscometer

Pharma and biopharma QC

Suspensions, syrups, gels, and biologics buffers all require viscosity control as part of specification testing. Report viscosity with the temperature and shear conditions that match the regulatory filing or SOP exactly. Changing method parameters between development and QC creates data that cannot be directly compared.

Food and beverage

Sensory profile and fill-line performance for sauces, beverages, and concentrates both depend on viscosity. Monitor batch-to-batch consistency and capture the impact of thermal processing steps on flow behavior. Shear thinning is common; report apparent viscosity at shear rates that represent actual process conditions.

Petrochemicals and lubricants

Kinematic viscosity at 40 °C and 100 °C is the input for viscosity index calculations and lube oil grading. Capillary methods under ISO and ASTM standards are the norm here. Temperature accuracy is particularly important because viscosity-temperature sensitivity is high for petroleum products.

Coatings, inks, and adhesives

Shear thinning behavior controls application, leveling, and sag resistance. Quote apparent viscosity at shear rates that simulate the relevant process step, whether that is blade coating, spraying, or roller application. A single low-shear viscosity number is not enough to predict coating performance.

Polymers and monomers

Solution viscosity or intrinsic viscosity in a defined solvent gives a cost-effective indicator of molecular weight trends and batch-to-batch consistency. For melt processing, rotational rheometry at elevated temperature characterizes processability. Accurate reagent purity matters here; small contaminants can shift intrinsic viscosity results. Merck's range of high quality chemistry products supports precise preparation of polymer solutions and calibration standards with traceable purity.

Bioprocess and diagnostics

Viscosity influences pumping efficiency, filtration throughput, and pipetting accuracy. Protein formulations often display non-Newtonian behavior at high concentration, so capillary methods alone are insufficient. Use rotational methods with appropriate geometry and report shear rate alongside every result.

Dynamic vs kinematic: which reporting format fits your decision?

Report in Kinematic Units (mm²/s, cSt)
  • Gravity-driven flow characterization
  • Petroleum product grading (ISO, ASTM)
  • Legacy standards requiring cSt
  • Newtonian liquids measured by capillary
Report in Dynamic Units (Pa·s, mPa·s)
  • Shear-controlled processing decisions
  • Non-Newtonian materials
  • Biologics and pharmaceutical formulations
  • When shear rate must be stated alongside the value

Technical summary and key takeaways on viscosity

Viscosity quantifies a fluid's resistance to flow. It is reported as dynamic viscosity in Pa·s or kinematic viscosity in mm²/s, always with the temperature stated, and for non-Newtonian materials with the shear rate included.

Method choice determines what the number actually means. Capillary methods give kinematic values for Newtonian fluids. Rotational methods give apparent viscosity and flow curves for complex materials. Standardize temperature control, run calibration verification across the full working range, and document every method parameter to keep results comparable across runs and sites.

Viscosity defines how a fluid resists motion. It is sensitive to temperature, shear, and measurement method. Select a technique that matches both the fluid type and the decision you need to make, then control temperature and report method details so results can be trusted by anyone who reads them.

Frequently asked questions about viscosity

What is the difference between dynamic and kinematic viscosity?
Dynamic viscosity relates shear stress to shear rate and is reported in Pa·s or mPa·s. Kinematic viscosity is dynamic viscosity divided by density, reported in mm²/s (cSt), and is typically measured with capillary viscometers. Both values must always be accompanied by the measurement temperature.
How do I choose between rotational and capillary viscometers?
Use a capillary viscometer for Newtonian liquids when you need kinematic viscosity values. Choose rotational for a wider viscosity range, non-Newtonian samples, or when you must report apparent viscosity at a defined shear rate. The nature of the sample and the specification format in your SOP should drive the decision.
How much does temperature affect viscosity readings?
A 1 °C shift can change viscosity by several percent for many liquids, and the effect is larger for more viscous materials. Control bath temperature tightly, allow full thermal equilibration at the sample point, and report the exact temperature setpoint with every result.
What common errors cause poor viscosity repeatability?
Typical causes include temperature drift, air bubbles in the sample, contamination from incomplete cleaning, and failing to state shear rate for non-Newtonian samples. Stabilize temperature before sampling, degas carefully, clean all wetted components fully between runs, and document all method parameters each time.
How should I report viscosity for non-Newtonian materials?
State apparent viscosity together with the method, geometry, temperature, and shear rate or spindle speed used. Note any time-dependent behavior such as thixotropy. Provide a full flow curve when the decision depends on performance across a range of process-relevant shear conditions.

Recommended Products

All products below are available through Chemical Express Thailand. Submit an inquiry for pricing, availability, or technical guidance.

Recommended Products Available at Chemical Express

We deliver a wide range of Measurement Fundamentals products from Merck and TCI and other international brands across Thailand. For any inquiry or purchase, fill out our Inquiry Form or chat with us instantly on Line.

Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.